To perform the test: • Caution: the fluid level rises
with higher temperatures
Place the calibrator on a flat surface with at least six inches of free and with the number and
space around the instrument. size of the probes placed into
the fluid.
Carefully insert the probe basket into the well and fill with the appro-
priate fluid . • Best results are obtained
with the probe inserted to
For optimal performance allow the manufacturer-recommended the full depth of the well.
warm-up period.
• The stabilization time of the
Insert the test probe to be calibrated into the well of the bath. For best Micro-Bath depends on the
performance, also insert a temperature standard for comparison. conditions and temperatures
involved. Typically stability is
Once the probe is inserted to the full depth of the bath, allow adequate achieved within ten minutes.
stabilization time for the test probe temperature to settle.
Additional resources
Once the probes have settled to the temperature of the bath, their indi-
For more in depth information
6 cation may be compared to the calibrator display temperature about this application check out
(or to a temperature standard such as a 1551AI. these videos and application notes
from Fluke.
r!!l Industrial Temperature
l!EJ Calibrators Workload Matrix
Process Calibration Tools:
Temperature Applications
980 www.tech-equipments.comlsales@tech-equipments.com
To perform the test: • Caution: the fluid level rises
with higher temperatures
Place the calibrator on a flat surface with at least six inches of free and with the number and
space around the instrument. size of the probes placed into
the fluid.
Carefully insert the probe basket into the well and fill with the appro-
priate fluid . • Best results are obtained
with the probe inserted to
For optimal performance allow the manufacturer-recommended the full depth of the well.
warm-up period.
• The stabilization time of the
Insert the test probe to be calibrated into the well of the bath. For best Micro-Bath depends on the
performance, also insert a temperature standard for comparison. conditions and temperatures
involved. Typically stability is
Once the probe is inserted to the full depth of the bath, allow adequate achieved within ten minutes.
stabilization time for the test probe temperature to settle.
Additional resources
Once the probes have settled to the temperature of the bath, their indi-
For more in depth information
6 cation may be compared to the calibrator display temperature about this application check out
(or to a temperature standard such as a 1551AI. these videos and application notes
from Fluke.