Kyoritsu KEW 6024PV PV Insulation Earth Tester
Zoom image

Kyoritsu KEW 6024PV PV Insulation Earth Tester


Not rated yet

Correct measuring for Insulation resistance even if the panel is under generating power

Waterproof design: Can measure under the bad weather condition

Memory function up to 1000 data

Luminescence button and Backlight

  • د.ك0.000KWD
Qty:  
In stock (1000 items available)
OR
Insulation resistancePV insulationInsulation
Earth resistance
Voltage measurement
General

Test voltage500V1000V250V500V1000V
Measuring range
(Auto range)
20.00/200.0/2000M/td>20.00/200.0/2000M/td>
Mid-scale value/td>50M/td>
Rated current/td>1.0-1.2mA
0.25M/td>0.5M/td>1M/td>
First effective
measuring range
1.51-200.0M/td>1.51-1000M/td>1.51-100.0M/td>1.51-200.0M/td>1.51-1000M/td>
Accuracy.5%rdgdgt.5%rdgdgt
Second effective
measuring range
0.00-1.50M/td>0.00-1.50M/td>1.20-1.50M/td>1.20-1.50M/td>1.20-1.50M/td>
200.1-2000M/td>1001-2000M/td>100.1
-2000M/td>
200.1
-2000M/td>
1001
-2000M/td>
Accuracy.0%rdgdgt
Open circuit voltage0 - +20%
Short circuit currentMax 1.5mA

Measuring range
(Auto range)
20.00/200.0/2000/td>
Accuracy.0%rdg.1(20ange) .0%rdgdgt (200/2000ange)

Measuring rangeAC 5-600V (45-65Hz)DC -000V
Accuracy.0%rdgdgt

Applicable StandardsIEC 61010-1 CAT 600V, CAT IV 300V Pollution2
IEC 61010-2-030, IEC 61010-031(EMC), IEC 60529_P54_
IEC 61557-1,-2,-5,-10, IEC 61326-1,2-2, EN50581(RoHS)
Power sourceLR6(AA)(1.5V) x 6
Dimensions /
Weight
84(L) x 184(W) x 133(D)mm / Approx. 900g (including batteries)
Accessories7196A(Test leads with remote control switch),
7244A(Test lead with alligater clip), 8017(Extension prod),
8072(CAT II standard prod),
8212-USB(USB adaptor with "KEW Report (Software)"),
9155(Shoulder strap), 9156(Carrying case), LR6x6, Instruction manual
Optional7243(L-shaped probe), 7245A(Precision measurement cord set),
8016(Hook type prod)