Technical Specification - Zerowaste® eMS
Gas Range Repeatability n
SF6 Purity 0-100% +/-0.5% FS'
Dewpoint -600( - 200( +/-0.5°( at -300( II
HF 0-200ppm +/-5% FS'
S02lo 0-150ppm +/-2% FS'
Gases S02 Hi 0-500ppm +/-2% FS'
Measured 0-4000ppm +/-2% FS'
and Options CF4lo 0-65% +/-1% FS'
CF4 Hi 0-1000ppm +/-5% FS'
CO 0-100ppm +/-5% FS'
H2S 0-250ppm 3% of FS'
R12 0-50% 1% of FS'
AIR
Operating Temperature Range -20"( to +500 ( (-4°F to +122°F)
FS·: Full Scale (of Measuring Range) Environmental Humidity 0-95% RH non-condensing
Continuo Conditions
OK 1 Physical Weight 15kg (33Ibs)
Specifications
Maximum l500 x W300 x 0470 mm
Dimensions (19.11 " x 12" x 18.5")
Gas Pressure 0.5 to 10 bar (7 to 145 PSI)
Measurement from 10-600 minutes
Intervals
Measurement Cycles 1-5
Battery Operating Time >8 hours
Charger Input 100-240V - SO/60Hz
Continuous Sa
Minutes 00 Minute
~
6
Zerowaste'" eMS Sampling Interva/Screen - 10 minutes minimum 8
Zerowaste'" eMS Sampling Intervo/ Screen - 600 minutes maximum
754 www.tech-equipments.comlsales@tech-equipments.com
Technical Specification - Zerowaste® eMS
Gas Range Repeatability n
SF6 Purity 0-100% +/-0.5% FS'
Dewpoint -600( - 200( +/-0.5°( at -300( II
HF 0-200ppm +/-5% FS'
S02lo 0-150ppm +/-2% FS'
Gases S02 Hi 0-500ppm +/-2% FS'
Measured 0-4000ppm +/-2% FS'
and Options CF4lo 0-65% +/-1% FS'
CF4 Hi 0-1000ppm +/-5% FS'
CO 0-100ppm +/-5% FS'
H2S 0-250ppm 3% of FS'
R12 0-50% 1% of FS'
AIR
Operating Temperature Range -20"( to +500 ( (-4°F to +122°F)
FS·: Full Scale (of Measuring Range) Environmental Humidity 0-95% RH non-condensing
Continuo Conditions
OK 1 Physical Weight 15kg (33Ibs)
Specifications
Maximum l500 x W300 x 0470 mm
Dimensions (19.11 " x 12" x 18.5")
Gas Pressure 0.5 to 10 bar (7 to 145 PSI)
Measurement from 10-600 minutes
Intervals
Measurement Cycles 1-5
Battery Operating Time >8 hours
Charger Input 100-240V - SO/60Hz
Continuous Sa
Minutes 00 Minute
~
6
Zerowaste'" eMS Sampling Interva/Screen - 10 minutes minimum 8
Zerowaste'" eMS Sampling Intervo/ Screen - 600 minutes maximum